Main features:
◆ Intelligent temperature and humidity and zero point compensation algorithm, 2 to 3 level above the target point full software automatic calibration function, to ensure the accuracy of the measurement results; Automatically detect the working status of each sensor and components at startup;
◆ Support PPM and mg/m3, %VOL and mg/L concentration units can be switched freely;
◆ Real-time data transmission, can accurately issue local and remote wireless alarm notifications (optional) in a variety of alarm modes;
◆ Automatic layout according to the number of sensors; Automatic storage, manual storage, single point storage, a variety of storage methods to choose from; The storage interval can be customized to store 30,1000 sets of data with date and time identification, such as optional micro SD memory card, which can store more than 232 million groups;
◆ Short-distance wireless transmission mode or GPRS wireless transmission mode with unlimited distance anywhere can transmit real-time detection data and alarm status to the security center (optional) through the built-in wireless module function;
◆ Support USB and RS1 serial port measurement data download; Support serial printing of micro portable printers.
Application fields:
2. Used for environmental protection environment monitoring such as air and pollution source emissions;
3. Used for petroleum, chemical, power station and other industrial process control, logging instrument supporting gas analysis and detection special;
4. Used in agriculture, medical and health care and scientific research experiments and other fields;
5. Gas content determination of various combustion tests in waste incineration landfill;
<>. Used for air monitoring in public places;
Introduction to the measurement principle:
◆ Infrared gas detection principleNon-dispersive
infrared spectroscopic gas concentration detection technologyNon-spectroscopic
infrared (NDIR) sensor is mostly used in the field of gas analysis as a means of spectral detection. It is mainly composed of infrared light source, air chamber, wavelength filter and infrared detector. The gas is introduced into the gas chamber by pumping, and the gas concentration is detected by electro-optical modulation relationship according to the absorption characteristics corresponding to the infrared spectrum of the gas at a specific wavelength.
Product parameters:
Product name |
Benchtop sulfur dioxide gas analyzer |
Detection of gases |
Sulfur dioxide (SO2) |
Detection principle |
Non-spectroscopic infrared (NDIR); Electrochemical principle optional |
Measuring range |
Non-spectroscopic infrared (NDIR) measurement range: 0-500ppm, 1000ppm optional; |
Resolution |
0.1 ppm(0~50ppm); 0.1ppm(0~100ppm); 1ppm(0~1000ppm); 1ppm (more than 1000ppm); |
Installation method |
(1) Bench type, rack type; (2) Wall mounting, with mounting bracket |
Indication error |
≤±3% FS (higher accuracy required depending on sensor performance) |
Reproducibility |
≤±1% |
Zero drift |
≤±1% (FS/year) |
Response time |
≤ 25 sec (T90), (higher requirements depending on sensor performance) |
Recovery time |
≤30 sec. (higher requirements depending on sensor performance) |
Warm-up time |
More than 120 seconds (depending on sensor configuration) |
Power supply |
220VAC +10%,-15% ; 50Hz ±5Hz |
fuse |
Rated current: 10A Size: 5mm*20mm |
Inlet air flow |
(0.5-0.8)L/min (flow fluctuation≤ 0.02L/min) |
Environmental stress |
86kPa~106kPa |
Ambient temperature |
-20°C~+55°C(typical) -40°C~+70°C(limit value) (higher temperature and humidity require optional pre-treatment system - FIX3200 gas online monitoring central treatment system) |
Ambient humidity |
10%~95%RH(non-condensing) |
Body material |
SPCC+ aluminum profile |
Output signal |
① Three-wire 4-20mA current signal output, which can be connected to various alarm controllers, PLC, DCS and other control systems |
Signal transmission method |
① Three-core (four-core) shielded cable transmission: up to 1000m to 2000m (single-core 1m2 shielded cable) |
Alarm mode |
External sound and light alarm, alarm sound: < 90 dB |
Alarm point settings |
A1 alarm value, A2 alarm value; TWA alarm value and STEL alarm value can be set separately |
display screen |
5 inch TFT touch TFT-LCD color LCD screen, 800x480 dot matrix LCD |
Mode of operation |
5 inch LCD touch screen touch operation + button auxiliary operation |
keystroke |
Main Interface + Lock Screen + Reset' |
Backlight |
Manual, automatic when alarming |
Direct reading |
Measured value, ambient temperature and humidity, date and time, maximum, minimum value |
Data logging |
Automatic, manual, single-point storage, 1-6 sensor storage intervals can be customized, and 300000 groups of data with date and time can be stored continuously. The optional micro-SD memory card can store more than 10 million groups of data |
Data download |
Support USB and RS232 serial port measurement data download |
Data printing |
Support serial printing of micro portable printer (optional). |
calibration |
Full software calibration function for level 2 to 3 target points, calibration calibration value can be set, and zero adjustment can be set with one button |
Calibrate standard gas requirements |
Zero point standard gas: 99.999% N2 (nitrogen); |
Explosion-proof rating |
Ex ia II.C T4 Ga (Intrinsically Safe) |
Degree of protection |
IP42 |
Operating mode |
Two operating modes: user mode and administrator mode are available |
size |
Standard 19-inch chassis, width 480mm× depth 310mm× height 130mm |
weight |
Approx. 7Kg |
Execution standards |
GB15322.1-2003,GB3836.1-2010,GB3836.2-2010,GB3836.4-2010 |
quality assurance |
The warranty period is one year, and the maintenance is free during the warranty period. |